Preventative Maintenance
General StatIM Preventative Maintenance:
Preventative Maintenance Schedule To ensure trouble-free performance, both the operator and the dealer must follow a preventative maintenance schedule.
NOTE: Please refer to your National, Regional, State or Safety laws for any additional reoccurring user testing that may be required.
Operator | ||
Daily | Water Reservoir | • Replace water as needed. • For opthalmic use, drain at the end of every workday, leave empty, and refill at the start of the next workday. |
Waste Bottle | • Empty the waste bottle every time you refill the water reservoir. • Fill the bottle with water, up to the MIN line marking. You may also add some chlorine-free disinfectant. | |
Weekly | Cassette | • Wash the interior of the cassette with dishwashing soap or a mild detergent that does not contain chlorine. • Scrub the inside with a cleaning pad designed for use with Teflon™-coated surfaces. • After removing all traces of the detergent, treat interior surfaces of the cassette with the STAT-DRI™Plus drying agent to enhance the drying process. Order more STAT-DRI™ Plus from your dealer quoting 2OZPLUS, 8OZPLUST, or 32OZPLUS |
Biological and/or Air Filter | • Check the filter for dirt and moisture. Replace if dirty. Call for service if wet. | |
Water Filter | • Check the water reservoir filter every week and clean if necessary. Replace only if necessary. | |
Every 6 months | Cassette Seal | • Replace every 500 cycles or six months (whichever is first), or whenever necessary. |
Biological and/or Air Filter | • Replace every 500 cycles or six months (whichever is first). |
Technician | ||
Once a year | Cassette | • Check the tray, lid and seal for damage. Replace if necessary. |
Biological Filter | • Inspect the biological filter for moisture. | |
Solenoid Valve | • Inspect the valve and clean if dirty. Replace the plunger if defective. | |
Pump | • Clean the filters, replace if dirty. | |
Check Valve | • Remove the exhaust tube from the back of the unit during the air drying phase. Check for air coming from the fitting. • Remove the air compressor tube from the check valve inlet while running a cycle. Make sure no steam is leaking from the valve. Replace if there are any leaks. | |
Water Reservoir | • Check the reservoir for dirt. Clean and rinse with steam process distilled water if necessary. | |
Calibration | • Calibrate the unit. |